DECEMBER 2000
WEEK ENDING DECEMBER 1
PRODUCT Mobile Digital Imaging Systems, C-Arm fluoroscopic x-rays: a) Series
7700; b) Compact 7700.
CODE None.
MANUFACTURER GE OEC Medical Systems, Inc., Salt
Lake City, Utah.
DISTRIBUTION Nationwide and international.
QUANTITY 218 units were distributed.
REASON The diagnostic x-ray devices were found
defective under 21 CFR 1003.10 and failed to comply with the performance
standard in 21 CFR 1020.32(c). The defect occurs as a result of the emission of
electronic product radiation that is unnecessary to the accomplishment of its
primary purpose and which creates a risk of injury to the operator and patient.
The system also failed to meet the fluoroscopic performance standard 21 CFR
1020.32(c) "X-ray production in the fluoroscopic mode shall be controlled
by a device which requires continuous pressure by the operator for the entire
time of any exposure".
PRODUCT Series 9800 Mobile Digital Imaging Systems, C-Arm Fluoroscopic x-rays.
CODE Series 9800.
MANUFACTURER GE OEC Medical Systems, Inc., Salt
Lake City, Utah.
DISTRIBUTION Nationwide and international.
QUANTITY 795 units were distributed.
REASON The diagnostic x-ray devices were found
defective under 21 CFR 1003.10. The defect occurs as a result of the emission
of electronic product radiation that is unnecessary to the accomplishment of
its primary purpose and which creates a risk of injury to the operator and
patient.
PRODUCT Vitros Calibration Diskette DRV 5285, for use with Vitros 250 and 950
Chemistry Systems, when used in conjunction with Generation 48 Digoxin Slides:
a) Catalog #825 1878 Vitros 250 Chemistry System; b) Catalog #871 6607 Vitros
950 Chemistry System.
CODE Data Release Version 5825.
MANUFACTURER Ortho-Clinical Diagnostics, Inc.,
Rochester, New York.
DISTRIBUTION Nationwide, England, France,
Germany, Italy.
QUANTITY 1,643 diskettes, affecting over 1,902
Digoxin Gen 48 packs.
REASON The Vitros Calibration Diskette DRV 5285
contained revised parameters that affected the time during which readings were
taken by the instrument to calculate reaction rates. This resulted in a shift
in digoxin predicted concentration. For customers who received Digoxin GEN 48
Slides and calibrated using an earlier release of the calibration diskette,
loading of DRV 5285 without recalibration of Digoxin GEN 48
slides resulted in a negative shift in digoxin
results by approximately 1.5 ng/ml.
PRODUCT Steri-Oss brand HL Cylindrical Implants (Endosseous Implants), TPS
Coated, 3.25D by14mm, Catalog #3214TPS, device that are surgically placed in
the upper or lower jaw to provide support for a prosthetic device, such as
artificial teeth and to restore the patients chewing function.
CODE Lot #993719.
MANUFACTURER Nobel Biocare USA, Inc., Yorba
Linda, California.
DISTRIBUTION Arizona, California, Connecticut,
Georgia, Kentucky, Massachusetts, New York, Oregon, Texas, Israel, Japan.
QUANTITY 59 units were distributed.
REASON The inner vial label and the vial cap
were mislabeled as 12 mm and not 14 mm.
PRODUCT META Pulse Generators.
CODE Model 1256.
MANUFACTURER Teletronics Pacing Systems, Miami
Lakes, Florida.
RECALLED BY St. Jude Medical, Cardiac Rhythm
Management Division, Sylmar, California
DISTRIBUTION Nationwide and international.
QUANTITY 1,848 were implanted.
REASON These devices have shown an increased
risk of failure of the pacer processing integrated circuits (IC) that controls
the pulse generator. This failure is most commonly characterized by sensing or
output anomalies, including the possibility of no output. Extensive analysis
has shown that the IC failure is due to electrostatic discharge (ESD) during
manufacturing.
PRODUCT GemStar Ambulatory PCA Infusion Pump.
CODE List #13000.
MANUFACTURER Abbott Laboratories, San Diego,
California.
DISTRIBUTION Nationwide and international.
QUANTITY Approximately 500 units were
distributed.
REASON The devices have a dimensional problem in
the latch retaining pin for the battery door that can work its way out of the
door and then the door cannot be closed.
PRODUCT Accu-Chek Voicemate System, Catalog #2030802, designed for testing
glucose in whole blood by visually impaired persons with diabetes.
CODE Lot #119164 and serial number on meter
7674342975.
MANUFACTURER SCI Systems, Inc., Huntsville,
Alabama.
RECALLED BY Roche Diagnostics Corporation,
Indianapolis, Indiana
DISTRIBUTION Nationwide.
QUANTITY 200 units were distributed.
REASON The Advantage blood glucose meter placed in
one system reports results in mmol/L instead of mg/Dl, which may result in an
18-fold difference in the numbers reported out to the patient over the voice
readout of the device.
PRODUCT Esprit Ventilator with 10.4-inch Display, color capable.
CODE Various serial numbers.
MANUFACTURER Respironics, Inc., Vista,
California.
DISTRIBUTION Nationwide and Japan.
QUANTITY 53 units were distributed.
REASON There is a design deficiency and the
backlight inverter PCB needs to be insulated.
PRODUCT Belisle brand Advanced Facial Muscular Therapy, an electrical muscle
stimulator system.
CODE All units.
MANUFACTURER Belisle Systems, Inc., Clearwater,
Florida.
DISTRIBUTION Texas, New York, California,
Kansas, Oklahoma, Georgia, Virginia, Oklahoma, Florida, North Dakota,
Louisiana.
QUANTITY 30 units.
REASON Device was marketed without a 510(k) or
PMA.
PRODUCT a) Ventrix(r) brand of Ventricular Tunneling Pressure Monitoring Kit,
Model NL950-V; b) Complete Ventricular Tunneling Kit, Model #NL950-C. Both
products are used to monitor intracranial pressure and drain cerebral spinal
fluid.
CODE Lot Numbers: a) 020400; b) VC1040 and
VC1041.
MANUFACTURER Integra Neurosciences, San Deigo,
California.
DISTRIBUTION Alabama, Arkansas, California,
Illinois, Massachusetts, North Carolina, New Jersey, New York, Ohio, Oregon,
Pennsylvania, Texas, Virginia, Venezuela.
QUANTITY 84 units were distributed.
REASON Some kits were packaged with an incorrect
trocar/sheath assembly. The kits were supposed to contain 9-10 French size, but
were mistakenly packaged with 4-6 French size.
PRODUCT Roche Control Serum N (Human), Catalog #0737119, intended as an
assayed quality control material to monitor the accuracy and precision at
normal concentration levels in quantitative clinical chemistry assays.
CODE Lot #U0434 EXP 1/31/02.
MANUFACTURER Roche Diagnostics Corporation,
Indianapolis, Indiana.
DISTRIBUTION Nationwide.
QUANTITY 750 kits with 20 vials each were
distributed.
REASON The normal control show a significant
decrease in recovery of Direct Bilirubin. The Direct Bilirubin recovery shows
bottle-to-bottle inconsistencies within the affected lot.
PRODUCT Roche Creatinine Enzymatic Reagent, COBAS Integra 400 and 700 Analyzer
Operators, Catalog #0763144, reagent intended for use on the COBAS Integra
analyzer for the kinetic quantitative determination of creatinine concentration
in serum, plasma or urine.
CODE All lot numbers.
MANUFACTURER Roche Diagnostics, Mannheim,
Germany.
RECALLED BY Roche Diagnostics, Corporation,
Indianapolis, Indiana
DISTRIBUTION Nationwide.
QUANTITY 2,010 units were distributed.
REASON Device may give falsely negative results
due to hemoglobin interference. The firm’s medical expert determined that this
negative interference can be clinically significant, especially in neonates and
children because their creatinine levels are lower in general.
PRODUCT Elecys Estradiol Assay, Catalog #1776002, an in-vitro quantitative
determination of estradiol in human serum and plasma.
CODE Lot #199871.
MANUFACTURER Roche Diagnostics, GmbH, Mannheim,
Germany.
RECALLED BY Roche Diagnostics Corporation,
Indianapolis, Indiana
DISTRIBUTION Nationwide.
QUANTITY 2,636 kits were distributed.
REASON A shift in recovery was identified for
this lot in comparison to the previous lot manufactured. This shift primarily
affects samples from males, post menopausal women and IVF patient.
PRODUCT Tina-Quant CRP, for the quantitative measure of C-reactive protein
(CRP) in serum and plasma on automated clinical chemistry analyzers: a) Catalog
#1551922; b) Catalog #1299859.
CODE All lot numbers.
MANUFACTURER Nissui Pharmaceutical Company,
Ltd., Tokyo, Japan.
RECALLED BY Roche Diagnostics Corporation,
Indianapolis, Indiana (distributor)
DISTRIBUTION Nationwide and Canada.
QUANTITY 3,042 kits were distributed.
REASON Neonatal specimens may initially produce
falsely elevated CRP results.
WEEK ENDING DECEMBER 8
PRODUCT AMK Single Step Patella Replacement Basket Instrument (26 mm), Catalog
#2500-29-000, an orthopaedic manual surgical instrument intended to be used to
cut or shape the patella bone to allow placement of the AMK patella prosthesis,
a component of the AMK total knee replacement system.
CODE Lot #U31EF1000.
MANUFACTURER DePuy Orthopaedics, Inc., Warsaw,
Indiana.
DISTRIBUTION Wisconsin.
QUANTITY 1 unit.
REASON The device was incorrectly labeled as
sterile when it was not properly sterilized.
PRODUCT Rapid Form Vacuum Immobilizer: a) Rapid Form Vacuum Immobilizer
2-piece set, stock #013490 containing 2 splints and 1 pump per set; b) Rapid
Form Vacuum Immobilizer 3-piece set, stock #013463 containing 3 splints and 1
pump per set; c) Rapid Form Pump item #013467.
CODE a) Lot LLF246 and LLG320; b) Lot LLF246; c)
None.
MANUFACTURER Cramer products, Inc., Gardner, Kansas.
DISTRIBUTION Nationwide.
QUANTITY a) 321 2-piece sets; b) 40 3-piece
sets; c) 22 replacement pumps were distributed.
REASON Some incorrect pumps were inadvertently
included in splint kits or sent out as replacement pumps.
PRODUCT Roche Creatinine Mira, Catalog #3033414, intended for use on the COBAS
MIRA analyzer for the kinetic quantitative determination of creatinine in serum
or urine.
CODE Lot #A12071 EXP 9/30/02.
MANUFACTURER Raichem, Division of Hemagen, San
Diego, California.
RECALLED BY Roche Diagnostics Corporation,
Indianapolis, Indiana
DISTRIBUTION Nationwide.
QUANTITY 2550 units.
REASON 1. The package insert, contained in the
kit, was printed with an incorrect reagent volume for COBAS MIRA analyzers
using software Version 8735.
2. The device may produce falsely elevated
creatinines dues to elevated trigliycerides in the serum samples.
PRODUCT a) Vitros 950 Chemistry System; b) Vitros 950AT Chemistry System.
CODE a) Serial Number less than 09501559; b)
Serial Number less than 09600129.
MANUFACTURER Ortho-Clinical Diagnostics, Inc.,
Rochester, New York.
DISTRIBUTION Nationwide and international.
QUANTITY a) 1,420; b) 30 units were distributed.
REASON An atypical pattern in relationship
between slide millivold used to calculate a predicted concentration and the
millivolt reading used to assess the slide impedance.
PRODUCT Immunosimplicity brand Is anti-ds DNA Test Kit (ONLY the Positive
control, lot #60600 is under recall, the kit cannot be used without the
positive control), Catalog #720700, an in-vitro diagnostic reagent kit.
CODE Lot #60600, EXP 6/01.Ortho-Clinical
Diagnostics, Inc., Rochester, New York.
MANUFACTURER Diamedix Corporation, Miami,
Florida.
DISTRIBUTION Tennessee, Florida, Pennsylvania,
Virginia, California, Ohio, Missouri, Georgia, Hawaii, Illinois, Costa Rica,
Italy.
QUANTITY 137 units were distributed.
REASON Failure of the Positive Control from this
lot number kit, to recover within its assigned labeled range.
PRODUCT Roche COBAS Integra Analyzer Operators: a) Cobas Integra 400, Catalog
#1045199;
b) COBAS Integra 700, Catalog #1044826; c) COBAS
Integra 700, Catalog #1046926.
CODE All serial numbers.
MANUFACTURER Tegimenta AG, Roche Diagnostics
Instrument Center, Rotkreuz, Switzerand.
RECALLED BY Roche Diagnostics Corporation,
Indianapolis, Indiana
DISTRIBUTION Nationwide.
QUANTITY Approximately 399 analyzers.
REASON Device reagent cassettes previously used
on a Cobas Integra analyzer can be interchanged on another Cobas Integra analyzer
and that analyzer will treat the used cassette as a new cassette. Under such
circumstances the instrument could sample the reagent at the wrong level
leading to erroneous results, and could incorrectly reconstitute and/or
dispense reagents.
PRODUCT Vascular prostheses: (a) C.R. Bard's Impra Flex(tm) Grafts with small
beading; (b) Distaflo(tm) Bypass Grafts.
CODE a) Catalog Nos. F1006TWS, F3006TWSC,
F310086TWS, F310086TWSC, F5006S, F5006SC, F5006TWS, F5006TWSC, F5008S,
F5008TWS, F5008TWSC, F6006TWS, F6006TWSC, F7004TWS, F7005TWS, F7005TWSC,
F7006S, F7006SC, F7006TWS, F7006TWSC, F7007TWS, F7007TWSC, F7008S, F7008SC,
F7008TWS, F7008TWSC, F70N75TSC, F70N75TWS, F70T7-4S, F70T74TSC, F70T74TWS,
F70T8-5S, F70T85TWS, F8006S, F8006SC, F8006TWS, F8006TWSC, F8007TWS, F8007TWSC,
F8008S, F8008SC, F8008TWS, F8008TWSC.
b) Catalog Nos. DF5006SC, DF5007SC, DF6006SC,
DF6007SC, DF7006SC, DF5007SC, DF8006SC, DF5007SC.
MANUFACTURER Impra, Inc., Tempe, Arizona.
DISTRIBUTION Canada, New Zealand, Korea, Brazil,
Australia, Peru, Japan, Belgium.
QUANTITY 4,270 units were distributed.
REASON The United Kingdom (UK) required the firm
to provide better instructions for use to caution the physicians when removing
the beading. The UK required the firm to perform a market withdrawal to resolve
this problem.
PRODUCT Olympus EVIS-Endoscopic Video Image and Data Systems for examination
of the gastrointestinal tract and colon:
a) Olympus Gastrointestinal Fiberscope GIF-2T20;
b) Olympus Gastrointestinal Videoscope
GIF-2T100.
CODE a) Olympus Gastrointestinal Fiberscope
Model GIF type 2T20 (GIF-2T20), Reprocessing Manual Part Number GR2180. Serial
Numbers of affected endoscopes: 2811017, 2811025, 2811028, 2811031, 2811034,
2811035.
b) Olympus Gastrointestinal Videoscope Model GIF
type 2T100 (GIF-2T100), Reprocessing Manual Part Number GR2180. Serial numbers
of affected endoscopes: 2861291-2861298, 2861300-2861304, 2861306-2861314,
2861316-2861337, 2861339-2861341, 2861344-2861356, 2861358-2861363,
2861365-2861379, 2861383-2861384, 2861386-2861396, 2861398-2861405, 2961406,
2961408-2961423, 2961425-2961426, 2961429-2961431, 2961433-2961437,
2961440-2961458, 2961460- 2961462, 2961464, 2961466-2961470, 2961472-2961492,
2961495-2961500, 2961502-2961516, 2961519.
MANUFACTURER Olympus Opto-Electronics, Ltd.
-Aizu Factory, Moden-machi, Fukushima-Ken Japan.
RECALLED BY Olympus America, Inc., Melville, New
York
DISTRIBUTION a) Alabama, Rhode Island,
Louisiana, Hawaii, Chile, Peru;b) Nationwide and international.
QUANTITY a) 6 units; b) 198 units were
distributed.
REASON The labeling instructions for
reprocessing both the two suction channel openings as well as the suction
switchover lever do not provide enough detailed information regarding
reprocessing of this feature and accessory.
WEEK ENDING DECEMBER 15
PRODUCT Sterichek Total Chlorine Reagent Strips, 100-test strips, Product
#811904, designed to measure low levels of total chlorine in feed water used to
prepare solutions for kidney dialysis.
CODE Product is sold both as a part of a kit and
as a single product. Kit lot numbers: 001069, 002069, and 003069. Kits expire
5/01. Bottle lots 002040, 004040, and 003040.
MANUFACTURER Environmental Test Systems,
Elkhart, Indiana.
DISTRIBUTION Nationwide and Japan.
QUANTITY Approximately 1,025 bottles.
REASON The test strips which are intended to
measure very low chlorine levels in water used for dialysis are ineffective
when used to measure high levels of chlorine.
PRODUCT Model SmoothLASE Alexandrite Laser System, used in dermatology.
CODE Undetermined.
MANUFACTURER Leisegang Medical, Inc., Boca
Raton, Florida.
DISTRIBUTION Nationwide.
QUANTITY 9 units were distributed.
REASON Device failed to comply with 21 CFR
1010.2, 1040.10(g), 1040.10(h), and 1040.11(a) in that the product lacked
certain required labels and operator's manual lacked calibration procedures and
labels.
PRODUCT Omnifit Normalized Hip Stem, for use in bipolar and total hip
arthroplasty in either a cemented or cementless : a) Omnifit Normalized Hip
Stem (Collarless), Catalog No. 6033-0625; b) Omnifit Normalized Hip Stem
(Collared), Catalog No. 6034-0625.
CODE a) Lot Code S00L303; b) Lot Code S00L269.
MANUFACTURER Howmedica Osteonics Corporation,
Allendale, New Jersey.
DISTRIBUTION California, Florida, Germany,
Korea, United Kingdom, Canada, Peru.
QUANTITY 22 units were distributed.
REASON Device was incorrectly labeled as sterile
when it was not properly sterilized.
PRODUCT VIP Gold Ventilator; b) VIP Sterling Ventilator.
CODE a) Catalog/Part #15653; b) Catalog/Part
#15654.
MANUFACTURER Bird Products Corporation, Palm
Springs, California.
DISTRIBUTION Nationwide and international.
QUANTITY 298 units were distributed.
REASON There is a failure to cycle and to alarm
during deliberate disconnect.
PRODUCT Electronic Video Colonoscope, a gastrointestinal video system: a) Model
EC-200LR, for the lower gastrointestinal tracts; b) Model No. 200LT, allows
examination and treatment from rectum to cecum.
CODE Serial Numbers: a) 4Cl6lDOO3, 2Cl6lDll8,
4Cl6lDOOl, 2Cl6lDll2, 2Cl6lDO96, 2Cl6lDlO3, 2Cl6lDlO6, 2Cl6lDlll, 2Cl6lDO97,
2CI61DO98, 2Cl6lDO99, 2Cl6lDll3, 4CI61DO05
b) 4C214DO16, 3C214DO35, 4C214DO06, 4C214DO20,
4C214DO03, 3C214DO27, 4C214DO29, 4C214DO13, 4C214DO09, 4C214DO10, 4C214DO11,
4C214DO08, 4C214DO27, 4C214DO30, 4C214DO04, 4C214DO21, 4C214DO22, 4C214DO26,
4C214DO07, 4C214DO25, 4C214DO02, 4C214DO12, 4C214DO01.
MANUFACTURER Fuji PhotoOptical Company, Ltd.,
Omiya City, Saitama, Japan.
RECALLED BY Fujinon, Inc., Wayne, New Jersey
DISTRIBUTION Nationwide and Canada.
QUANTITY 36 units.
REASON The video colnonoscope bending sections
cannot be properly manipulated.
PRODUCT Vitros C-Reactive Protein (CRP) Slides GEN 18, quantitatively measure
C-reactive protein concentration in serum and plasma.
CODE 3718-0132-0111 EXP 9/1/99.
MANUFACTURER Ortho-Clinical Diagnostics, Inc.,
Rochester, New York.
DISTRIBUTION Germany and Sweden.
QUANTITY 291 packs were distributed.
REASON Slides contained cartridges from two
different generations that were mixed.
PRODUCT Immulite brand of CMV IgG Test Kit, Model L2KCV2, an in-vitro
diagnostic device.
CODE Lot #102A.
MANUFACTURER Diagnostic Products Corporation,
Los Angeles, California.
DISTRIBUTION Kansas, New Jersey, India, The
Netherlands.
QUANTITY 7 kits were distributed.
REASON The barcode used to label the adjustor
kit component had an error that caused the system to reject it. The printing
error prevents the barcode from being read and the instruments will not process
the assay and patient results will not be obtained.
PRODUCT ACN Biopsy Needle, Part #(model) ACN1410MF, used for obtaining
multiple core samples from soft tissue such as the liver, kidney, prostate,
breast and various other soft tissue lesions.
CODE PART NUMBER ACN1410MF and the following lot
numbers: 01101469, 01101470, 01101471, 01221644, 01231645, 01241646, 01241647,
01241648, 01261800, 01291801, 01381715, 01381737, 01391736, 01441738, 01581917,
01581961, 01602025, 01612157, 01682158, 01882602, 01922675, 01952765, 02002809,
02022810, 02032811, 02062812, 02203477, 02283673, 02303674, 02343877, 02373933,
02373934, 02494027, 02564117, 02624228, 02634227 and 02664354.
MANUFACTURER Medical Device Technologies, Inc.,
Gainesville, Florida.
DISTRIBUTION Nationwide and international.
QUANTITY 2,374 units were distributed.
REASON Needle did not operate smoothly in gun,
potentially affecting QUANTITY of sample.
PRODUCT a) Osteonics C-Taper Head 26 mm, + 0, Catalog No. 06-2600; b)
Osteonics C-Taper Head 26 mm, +10, Catalog No. 06-2610, indicated for use of
the Osteonics Ion Implanted Femoral Bearing Series as part of a universal hip
replacement.
CODE a) Lot No. 56629202 b) Lot No. 56584201.
MANUFACTURER Howmedica Osteonics Corporation,
Allendale, New Jersey.
DISTRIBUTION Nationwide and Australia.
QUANTITY 32 units were distributed.
REASON Packages labeled Catalog No. 06-2600
(Osteonics C-Taper Head 26 mm, + 0) contained Catalog No. 06-2610 (Osteonics
C-Taper Head 26 mm, +10), and vice versa.
PRODUCT Immulite 2000 PAP IVD Test Kit, for the chemiluminescent enzyme
immunometric assay for the quantitative measurement of prostatic acid
phosphatase.
CODE Lot # L2KPA2 6103 EXP 11/30/00.
MANUFACTURER Diagnostic Products Corporation,
Los Angeles, California.
DISTRIBUTION Arizona, California, Missouri, New
Jersey, Texas, Utah, Australia, Brazil, Germany, Italy, Spain, Sweden,
Switzerland, Turkey, United Kingdom.
QUANTITY 17 kits were distributed.
REASON The second full reagent wedge of the same
lot on the system will erroneously be recognized as being empty, no further
processing will take place and test results will not be possible to obtain.
PRODUCT ROCHE c.f.a.s. Calibrator for automated systems, Catalog No. 759350,
used for calibration of Roche methods on clinical chemistry analyzers.
CODE Lot Nos. 199608 through 199611, 199845
through 199848.
MANUFACTURER Roche Diagnostics, GmbH, Mannheim,
Germany.
RECALLED BY Roche Diagnostics Corp.,
Indianapolis, Indiana
DISTRIBUTION Nationwide and Puerto Rico.
QUANTITY Approximately 10,000 kits were
distributed.
REASON The setpoints of the master lot and
customer lot produce small bias that may produce calcium results too high.
PRODUCT Abbott Multiconstituent Calibrator, for use with the ALCYON Analyzer and
AEROSET System for the calibration of albumin, calcium, cholesterol,
creatinine, glucose, phosphorus, total protein, triglycerides, urea nitrogen,
and uric acid.
CODE List #1E65-01 and Lot #54650M100.
MANUFACTURER Medical Analysis Systems, Inc.,
Camarillo, California.
RECALLED BY Abbott Laboratories, Inc., Irving,
Texas
DISTRIBUTION Nationwide, Mexico, Venezuela,
Colombia, Dominican Republic, Canada, Germany, Hong Kong, Singapore, Australia,
New Zealand, Curaco..
QUANTITY 1,733 kits were distributed.
REASON The stability failure of the
multiconstituent calibrator (MCC). Creatinine concentration has decreased from
the test June 1, 2000 due to high iron content. The decrease in creatinine
concentration in the MCC may result in an upward shift of creatinine results of
up to 12%. The expiration date of this lot is November 3, 2000.
WEEK ENDING DECEMBER 29
PRODUCT FastTake Compact Blood
Glucose Monitoring System, used to quantitatively measure glucose (sugar)
levels in whole blood taken during home-care use, under the following trade
names:One Touch FastTake Compact Blood Glucose Monitoring System (in the U.S.
and Canada); PocketScan Compact Blood Glucose Monitoring System (in the United
Kingdom); EuroFlash Compact Blood Glucose Monitoring System (in Europe);
SmartScan Compact Blood Glucose Monitoring System (in Asia, Middle East,
Africa, Europe).
CODE All meters with serial numbers starting with K, L, M
and N.
MANUFACTURER Inverness Medical, Inc., Waltham,
Massachusetts.
RECALLED BY LifeScan, Inc., Milpitas, California
DISTRIBUTION Nationwide.
QUANTITY 846,874 meters were distributed.
REASON The meter may display a "y" character
instead of a number in the test result.
PRODUCT VITROS Chemistry Slides: (1)
Albumin chemistry slides - labeled "ALB Slides", for quantitative
measurement of albumin concentrations in serum or plasma (Catalog #819 6057);
(2)
BUN/Urea Chemistry Slides, labeled "BUN slides", for quantitative
measurement of urea nitrogen concentration in serum, plasma and urine (Catalog
#174 3418);
(3)
Glucose Chemistry Slides, labeled "GLU Slides", for quantitative
measurement of glucose concentrations in serum, plasma, urine and CSF (Catalog
#170 7801);
(4)
Potassium Chemistry Slides, labeled "K+ Slides", for the quantitative
measurement of
potassium
concentrations in serum, plasma, and urine (Catalog #161 6200).
CODE a) 0914-0169-1087, 0914-0169-1089, 0913-0170-1078;
b) 0111-0503-1067; c) 0001-0543-1017, xxx1018, xxx1020, xxx1028, xxx1029, and
xxx1327; d) 4102-0394-1038, xxx1104, and xxx1106.
MANUFACTURER Ortho-Clinical Diagnostics, Inc., Rochester,
New York.
DISTRIBUTION Nationwide.
QUANTITY 3,846 packs were distributed.
REASON The moisture content (humidity) inside the cartridge
sleeve might be lower than is typical for these products. The Slides are conditioned at a defined temperature and
relative humidity before packaging to optimize performance and handling. The effect of this deviation on shelf
life and on-board stability are unknown.
Stability claims may not be met and performance may not be effective for
intended use.
PRODUCT Dall-Miles Cable System,
Dall-Miles Beaded Cable Sleeve Set, Wire Surgical, used for spinal wiring,
sternotomy applications, cerclage procedures, and trauma surgery of the
shoulder, elbow, knee, hip, or ankle. Product is gamma sterilized.
CODE Lot Numbers: IGTW01, IGTX01, IGTZ01.
MANUFACTURER SAVA Industries, Inc., Riverdale, New Jersey.
RECALLED BY Howmedica Osteonics Corp., Rutherford, New
Jersey
DISTRIBUTION Alabama, California, Connecticut, Florida,
Kentucky, Massachusetts, Minnesota, Montana, New York, Oregon, Tennessee,
Texas, Utah, Idaho, Indiana, Maryland, Virginia, Germany, England, Switzerland,
Spain, Auckland.
QUANTITY 249 units were distributed.
REASON The packages labeled 2.0 beaded cable actually
contained non-beaded cable.
PRODUCT Bayer ADVIA 120 Automated
Hematology Systems:Advia 120 Automated Hematology System;
a)
Technicon H1 Hematology Systems;
b)
Technicon H2 Hematology Systems;
c)
Technicon H3 Hematology Systems;
d)
Technicon H1 Jr Hematology Systems;
e)
Technicon H1E Hematology Systems.
CODE All serial numbers.
MANUFACTURER Bayer Diagnostics Mfg., Swords, Co Dublin,
Ireland.
RECALLED BY Bayer Corporation, Elkhart, Indiana
(distributor)
DISTRIBUTION Nationwide and international.
QUANTITY 1,800 units were distributed.
REASON A customer was sprayed with the contents of the
waste tank, since he did not release the pressure before opening